Oxide and Doubly-charged Species: Measured under identical operating conditions used to achieve sensitivity and detection-limit specifications.
CeO+/Ce+
< 0.025
Ce++/Ce+
< 0.03
Background Signal:Mass 220 < 1 cps
Short-term Precision:< 3% RSD
Long-term Stability:< 4% RSD over four hours
Isotope-ratio Precision:< 0.08*% RSD (*or within a factor of two of the counting statistics limit)
Mass Calibration Stability:< 0.05 amu over eight hours of continuous operation
Quadrupole Peak Hop (Slew) Speed:1.6 M amu/sec
Quadrupole Scan Speed:5000 amu/sec
Abundance Sensitivity:Measured at 238U Better than 1.0 x 10-6 at low mass side of peak Better than 1.0 x 10-7 at high mass side of peak
Detector Linear Range:The SimulScan? detection system operates from < 0.1 cps to > 109 cps. This provides over 10 orders of magnitude of linear dynamic range in a single continuous scan.
Transient Data Acquisition Speed:> 3000 temporal data points/sec maximum